Triple Echo Steady State (TESS) Relaxometry

Rahel Heule1, Carl Ganter2, Oliver Bieri1

1Department of Radiology, University of Basel Hospital, Basel, Switzerland; 2Department of Radiology, Klinikum rechts der Isar, Technische Universitt Mnchen, Munich, Germany

A variety of SSFP methods have been proposed so far and are known in the literature for fast relaxometry, but all of them are sensitive to B1, and show some more or less pronounced mixed T2/T1 sensitivity. In this work, a new rapid relaxometry method is introduced based on a triple echo steady state (TESS) approach. Relaxometry with TESS is not biased by T2/T1, is insensitive to B0 heterogeneities, and, surprisingly, for T2 not affected by B1 field errors. As a result, the new proposed method is of high interest for fast and reliable relaxometry in the clinical routine.

Keywords

according accurate addition affected allowing allows analytical applicability articular attracted audience axial bias biased brackets calibration cartilage certain clinicians column combination commonly completely confounding considerably constraints contrast contributions controls correspondence definitions dependencies derived development diagram doped double echoes either entail errors especially estimation evaluated excellent excitation expected expressions extraordinary falls fast feature field fields find free golden guess heterogeneities hospital illustrated imperfect in vivo independent indicated influence insensitive instead internal investigated iterations joint knee least leftmost limited lowest manganese mapping maps middle might mode modes muscle namely nine nominal none nonlinear offers optimized overestimated overestimation perfect procedure prominent prone pronounced proposed pulses quantification quantify radiology rapid recalculation refer refocusing related requiring resolution revealed rightmost scientists search section selected sensitivity serve simulations simultaneously spacing spatial special spectral spin squares starting steady stimulated studies summarized surprisingly system table target third threshold tissue transmit triple tubes turn typically ultra unaffected updated variations vitro whereas whole yellow