Abstract #0544

In Vivo Measurement of T1 & T2 Relaxation Times in Healthy Human Prostate at 7T

MAGNA25Marnix C. Maas1, Miriam W. Lagemaat2, Stephan Orzada3, 4, Thiele Kobus2, Oliver Kraff3, Stefan Maderwald3, Andreas K. Bitz3, 4, Mark E. Ladd3, 4, Tom W.J. Scheenen2, 3

1Radiology , Radboud University Nijmegen Medical Centre, Nijmegen, Netherlands; 2Radiology, Radboud University Nijmegen Medical Centre, Nijmegen, GLD, Netherlands; 3Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Duisburg-Essen, Essen, Germany; 4Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany

Knowledge of T1 and T2 relaxation times can aid the development and optimization of dedicated pulse sequences for imaging and spectroscopy at ultra-high field (≥7T). We have measured these quantities in human prostate and surrounding tissues at 7T in healthy volunteers using progressive saturation gradient echo imaging with fixed TR and varying flip angle for T1 and multiple-contrast spin-echo imaging for T2. The results suggested a TE of 6070 ms for optimum contrast between the prostate peripheral zone and transition zones in fast spin-echo imaging, which was indeed shown to yield contrast between these tissues.

Keywords

abdominal absorption accounting achieved adding adiabatic agreement amplitude anatomic anatomical approx approximately approximation around array assumed attenuation available bandwidth bias body calibration cancer challenging clear coil complete considering constants constraints contour contrast contribution corrected correspond created crudely curve decay development difficult drawn durations echoes elements ensure equipment examined example excitation experiment exponential field gradient healthy help human improvement in vivo indicated influence inhomogeneity inhomogeneous latter least length linear linearized magnitude maps meander measured median medical minimize minutes motion movement muscle muscles must oliver operate optimization optimized optimizing optimum organ particularly peak peripheral peristaltic pixel power progressive prostate pulse quantify quantitative quantities radiology ranks rectal reduce refocusing repeated requiring resolution robustness saturated saturation scaled sensitivity shimming signed simulations slice smooth software source spacing specified spectroscopy spin spoiled squares strength strengths strongly substantial supervision surrounding system table tissues train transformed transition uncertainties uncertainty useful variation varied varying volunteer volunteers water whole yield yielded zone zones