Meeting Banner
Abstract #0543

Design & Evaluation of an Eight Channel TX/RX Hybrid Applicator for Imaging & Targeted RF-Heating at 7.0 T

MAGNA25Lukas Winter1, Celal zerdem1, Werner Hoffmann2, Helmar Waiczies1, Davide Santoro1, Alexander Mller1, Tomasz Lindel1, Wolfgang Renz1, 3, Conrad Martin1, Frank Seifert2, Bernd Ittermann2, Thoralf Niendorf1, 4

1Berlin Ultrahigh Field Facility (B.U.F.F.), Max-Delbrueck-Center for Molecular Medicine, Berlin, Germany; 2Physikalisch Technische Bundesanstalt (PTB), Berlin, Germany; 3Siemens Healthcare, Erlangen, Germany; 4Experimental and Clinical Research Center (ECRC), Charit - University Medicine Campus Berlin Buch, Berlin, Germany

Combining RF hyperthermia and MR imaging is conceptually appealing to pursue spatially and temporally controlled and monitored RF heating. The benefits of this approach could be used as an adjunctive therapy for established cancer treatments including radiotherapy and chemotherapy , targeted drug delivery and targeted MR contrast agent delivery. This study demonstrates the feasibility of an 8 channel TX/RX hybrid applicator tailored for imaging and targeted hyperthermia at 7.0T

Keywords

abdominal allow amplifier amplitude antenna antennas anticipate appealing applications applicator applied approximately beyond biological bolus brain built campus cancer capabilities channel chemotherapy combining commonly components controlled cool coronal correlation cycle decoupling deep delivery depicted design designed diameter dimensions dipole distributions dominate duty eight electric element elements encouraging evaluation exceeds experimental explore family feasibility field fields focal frank frequency goes gradient healthy heating heavy help highly hybrid implications improvements in vivo individual insert integrated intrinsic lancet liver local losses maps martin matching matrix measured medicine modulated modulation molecular move nearly noise numerical optical performance phantom phys power preliminary prevent promise proof proposed proton pulse pursue radiative radiotherapy radius receive recent rectangular reduce reports sensors settings setup sharpen shorten simulated simulation simulations since spatially spin steering studies studio successfully surface symmetrical system targeted technology temperature therapy thermo together towards traits transient transmit treatment treatments tuning ultimate ultrahigh validated variable view virtual water whole width yang