Optimized Three Dimensional Sodium Imaging of the Human Heart on a Clinical 3T Scanner

Neville D. Gai1, Carlos Rochitte2, Marcelo Nacif3, David A. Bluemke1

1Radiology & Imaging Sciences, National Institutes of Health, Bethesda, MD, United States; 2Heart Institute - InCor - University of So Paulo Medical School, So Paulo, SP, Brazil; 3Radiology Department of Faculdade de Medicina, Universidade Federal Fluminense, Niteri, RJ, Brazil

Sodium imaging provides a direct window into cell viability. In the heart, tissue viability after an ischemic event is an important determinant in the decision to intervene with corrective measures. Sodium MRI of the heart is challenging due to extremely poor SNR and motion artifacts. In this work, we explore a 3D stack of spirals trajectory for imaging the entire human heart on a clinical 3T scanner. Optimization of sequence parameters based on Bloch simulations (and factors affecting the point spread function) makes sodium imaging the entire heart possible within a reasonable scan time (< 8 min).

Keywords

absence accounts acquiring acquisition allows artifact artifacts away axis background begins beneficial best blurring broadband cardiac challenging chief choice clear clearer clinical clinically coil combination combinations conditions conservative considerations constant continuous counter coverage definition described dimensional displacement done effective efficient entire equations especially even exact exhibited extended fall fast feasibility federal field final finally five gradient greatly health heart hence human ideal improved in vivo inconclusive increasing inhomogeneity institutes integrated length lengthier long longer lost materials measured medical minutes motion movement myocardium negating noise nutation object optimization optimized parts patient phantom phys picture position possibility preprocessed prescriptions pulse radiology rapid reasonable receive reduced reformatted release requires resolution resonant scanner scanners scanning school sciences section segmented septum several short simulations since slice slices sodium software solution space species spent spiral spirals spoiled spoiling still studies sufficient suitable sung surface train trajectory transmit traversing triggered typical typically variation ventricle version versions volume volunteers whether window