Chemical Shift Correction in Fat-Water Separation Using Two-Point Dixon SSFP

Yi Wang1, 2, Glen Morrell2, Dennis L. Parker2

1Neurology, University of California, Los Angeles, Los Angeles, CA, United States; 2Utah Center for Advanced Imaging Research, University of Utah, Salt Lake City, UT, United States

SSFP sequences offer superior signal intensity in a relative short time. The potential of a dual-echo SSFP sequence has been investigated for breast imaging applications, where fat signal is separated from the water signal using the two-point Dixon technique [3]. Dual echo SSFP provides robust water-only and fat-only images in the presence of B0 and B1 inhomogeneity. Due to the sensitivity of SSFP to off resonance, short TR is needed to minimize the banding artifact. The two echoes required for the two-point Dixon fat-water separation can be acquired using a bipolar gradient. Compared to mono-polar readout, a bipolar gradient can reduce echo spacing and therefore result in more efficient readout. However, due to the opposite chemical shift, mis-registration between images from the two echoes using bipolar readouts exists. In this work, we improved the dual-echo 3D SSFP sequence by correcting the bi-directional chemical shift error in two-point Dixon fat/water separation.

Keywords

ability acknowledgments acquisition addition adjacent advanced alternating applications arrows artifact background banding bipolar black blue bottle bottles bottom breast care caused chemical city clearly combined composed consisting correcting correction correctly design diagram directional displayed dual echoes edges efficient error examined example exists fashion fixed focused foundation free glen gradient gradients grant health illustrated implemented implies improved in vivo incorrectly indicated induced inhomogeneity intensity interleaved intervals intrinsic investigated iris lake leading location matrix minimize mono needed negative neurology offer opposite parker performance phantom polar polarities polarity pork positive potential precession presence pulse pulses pure readout readouts reconstruct reconstructed reduce registration required respectively robust salt sample scanner schematic selective sensitivity separated separation short slice slices source spacing steady superior supported suppression surgery trio ultrasound unique visible volume water