Meeting Banner
Abstract #2819

The New EU Proposal Vs. 2004/40/EC - An MR Exposure Data Comparison

Jens Groebner1, 2, Michael Bock1

1Dept. of Radiology / Medical Physics, University Medical Center Freiburg, Freiburg, BW, Germany; 2Dept. of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, BW, Germany

The EU announced that the directive 2004/40/EC will be replaced by a proposal of the European Commission (EC). In this work the EC limit values and the 2004/40/ECs limit values are compared to MR exposure data. 37 MR workers were asked to wear a field probe in the magnet room (up to 7T). Results show that the new limits (both peak and average) can be exceeded during routine work. With an optic/acoustic response system the ECs proposal can help reducing transient effects like vertigo or magnetophosphenes.

Keywords

according acoustic additional allows always amounted approaches asked black blue bock calibrated cancer caused coil considered dash dashed dataset datasets density depicts dept detected diameter directive dotted draft electro exceed exceeded exceeds exposure feedback field fields flux forehead frequency future hand help horizontal like limit limits mater materials measured medical movements nausea near optic peak physics planned probe procedure procedures proposal proposals proposed radiology recorded reducing regulating replace replaced represent response restrictive rotation rotations routine sets simultaneously since stay suggested summarizes system systems table trans transient translation translations trio union varying vertigo volunteer volunteers wear workers