Abstract #2818

Simultaneous PET/MR with Continuous Table Motion: The Effect of Table Motion Speed on Image Quality

Harald Braun1, Susanne Ziegler1, Harald H. Quick1

1Institute of Medical Physics, University of Erlangen-Nrnberg, Erlangen, Germany

In simultaneous whole-body PET/MR imaging, data is usually acquired in a multi-station (step-and-shoot) approach. In contrast, an acquisition paradigm with continuous table motion that was introduced recently allows for seamless whole-body coverage and greater flexibility in the planning workflow. To allow the use of continuous table motion acquisitions in clinical protocols, it must be evaluated which impact table motion speed has on MR and PET image quality. This was performed for the available MR protocols that support continuous table motion (FLASH 2D, FLASH 3D, TSE, HASTE, BLADE) and for PET imaging for several different table speeds.

Keywords

accuracy achieve achieved acquisition acquisitions addition advantages affected allow allowing allows appeared around arrows artifact artifacts assess attributed audience axial axis balanced behavior best biograph biomedical blade body bottom built certain clinical clinically command consistently constrained contains continuous control coronal correction coverage crisper custom cylindrical decrease decreased delineation develop diameter distortions encoding evaluated exhibit feasible field flash fluid fluke format fraction geometric geometry greatly hardly hardware haste hybrid impact important inherent inherently initiated innovative insert installed introduced iterations length like manually measured measures medical mode model modified molecular mostly motion moving must necessary noticeable noticeably patients phantom phantoms photography phys physicians physics pipeline place poster processing protocols providing quality quick randoms recently reconstructed reconstruction reduced resolution respect retaining scanner seamless seem served several simplified simultaneous slice slices slight slightly spatial speed speeds starting station stop structured studies subsets superior support system table taking target tools traditional various varying volumes water whole