16-Channel Receive Array Insert for Breast Imaging at 7T

Samantha By1, Joseph Rispoli1, Jiaming Cui2, Sergey Cheshkov3, 4, Ivan E. Dimitrov3, 5, Craig R. Malloy3, 4, Steve Wright1, 2, Mary P. McDougall1, 2

1Biomedical Engineering, Texas A&M University, College Station, TX, United States; 2Electrical and Computer Engineering, Texas A&M University, College Station, TX, United States; 3Advanced Imaging Research Center, University of Texas Southwestern Medical Center, Dallas, TX, United States; 4Radiology, University of Texas Southwestern Medical Center, Dallas, TX, United States; 5Philips Medical Systems, Cleveland, OH, United States

B1 inhomogeneities associated with high fields make effective imaging and hardware design challenging. "Forced current excitation" (FCE) provides homogeneous excitation and ease of tuning, irrespective of asymmetric loading, but the necessary incorporation of transmission line lengths in the resonant structure decreases SNR. This work presents the combined use of a FCE transmit breast coil with a 16-channel receive array insert to increase the receive sensitivity while taking advantage of the benefits of FCE during transmission. Breast imaging results indicate an improvement in SNR of up to six times from the insert as compared to the FCE coil in T/R mode.

Keywords

acknowledge acknowledgment activation active address advanced allows approved array authors away ball benefits biomedical blocking body bottom breast cancer channel circuits clad coil college combined common computer conditions configuration connecting contract control copper creating custom customized decoupling decreases depicted described designed despite detail detuning diagnosis diagram diode disadvantage discusses effectively electrical element elements enable engineering enhanced equal excitation fabricated feed field fields force forced gratefully hexagon housing human impedance improved improvement in vivo inclusion independent indicates induced inhomogeneity insert issue keep known larger lengths loading loss medical metabolites minimize mitigates mode monitoring moving mutual nearest neighbors noise nominal onto pack passive pentagon photograph place pose potential profiles promises protocol quantification radiology receive receiver reception rendering reported resonant respective scanner sensitivity several shell short side sided simplified soccer southwestern spectroscopy station still straightforward structure substantial support suppression systems taking thick transmission transmit traps tuning unequal useful utilizing various view volume volunteer wavelengths whole wright