Switchable Bilateral/Unilateral 7T Breast Coil Using Forced Current Excitation

Jiaming Cui1, Ivan E. Dimitrov2, 3, Sergey Cheshkov2, 4, Mary P. McDougall, 15, Craig R. Malloy2, 4, Steven M. Wright1, 5

1Electrical and Computer Engineering, Texas A&M University, College Station, TX, United States; 2Advanced Imaging Research Center, University of Texas Southwestern Medical Center, Dallas, TX, United States; 3Philips Medical systems, Cleveland, OH, United States; 4Department of Radiology, University of Texas Southwestern Medical Center, Dallas, TX, United States; 5Biomedical Engineering, Texas A&M University, College Station, TX, United States

Breast imaging and spectroscopy are of interest at 7T. In this abstract, a 7T Forced Current Excited (FCE) bilateral breast coil, which can be electronically switched to unilateral excitation mode, was developed. The coil was designed to provide the large field-of-view needed for bilateral breast imaging, while allowing unilateral mode when higher excitation fields are needed. Extending the forced current approach to two Helmholtz pairs avoided any difficulties with mode splits due to coupling between the closely spaced coils. The coil was tested at a 7T scanner using phantoms, demonstrating highly effective mode switching and excellent homogeneity.

Keywords

achieve acquisitions additionally advanced advantages allowing among array avoid avoided axis biasing bilateral bilaterally biomedical board body breast breasts circle circuitry circuits closely coexistence coil coils college common complicate computer conditions connected contra convenient coronal coupling create created dashed decoupled decouples desirable desired developed diagnostic diagram difficulty diode disables disconnects driven dual effectively electrical electronically elements eliminates enable enabled enables engineering ensures equal excellent excitation excite expected field fold forced form forward frequency future half highest homogeneity identical implemented individual individually inhomogeneity inserts integrated invest lateral latter loading loops many marked match medical mode modes must near network occur open otherwise output overcome pair pairs phantom phantoms polarity potential power prevents proved radiology receive resolution saved scanner sensitive separation share shunt simulate southwestern spaced spectroscopy splitting station successfully supplies suppression switchable switched switching system systems taken target throughout together transmit tune tuning turn uniform unilateral unwanted upgraded view voltage whole wright