Abstract #2573

Field Probe Based Shot-To-Shot B0 Correction for Multi-Shot Breast-DWI at 7T

Tijl A. van der Velden1, Vincent Oltman Boer1, Peter R. Luijten1, Dennis W.J. Klomp1

1Radiology, University Medical Center Utrecht, Utrecht, Netherlands

Field probes have been used for shot-to-shot B0 correction multi shot DWI images of the human breast at 7T. B0 correction is necessary for multi shot imaging to prevent ghosting due to temporal B0 fluctuations caused by physiological motions. In this study we demonstrate that the information simultaneously obtained from field probes reduce ghosting artefacts in multi-shot DWI at 7T in the human breast.

Keywords

achieve acquisition already applied around becomes benign blurring boundaries breast built calibrated cause causes clinical close coherence coil coils combination considering correct corrected correction detection diffusion directly distinguish distortions dual dynamic effective equipped especially every feasible field fields fluctuations full ghost ghosting gradient healthy home human improved improving inherent inside instability inter interfaced just leads lesions limited lipid local malignant measure measured medical minimal monitoring motion moving particularly peter physiological placed probe probes prone quadrature quality radiology real recently reduce reduced reduces reducing regularly removing resolution resolutions respiratory restored retrospective reversal scanned selection sensitivity severe short shot shots slice space spatial spin stored substantially suffers suppression susceptibility system thereby tuned unilateral variability variation volunteer water weighed worse zero