Phase Encoding Correction for 3D FSE Microscopy

Jonathan Bishop1, R Mark Henkelman1, 2, Brian J. Nieman1, 2

1Hospital for Sick Children, Toronto, ON, Canada; 2Medical Biophysics, University of Toronto, Toronto, ON, Canada

Phase errors generated by the phase encoding gradients of a 3D fast spin-echo sequence for preclinical microscopy are measured with a novel prescan sequence and applied as a retrospective correction at reconstruction to improve image quality.

Keywords

acquisition additional addresses although amplitude applied applies array artifact artifacts assigned axes axis biophysics bishop bottom brain children circle clinical coarse coil compensation composed conducted context contributed corrected correction cross cylindrical desirable dimension dimensions done echoes effectively element employ encode encodes encoding enhanced equally error errors even extends fast feathered fitted fixed full generate generated gradient gradients hospital imperfections implementation implemented include individually interpolated isotropic just lack latter locations logical mark matrix measure measured measures medical microscopy moment mouse novel omit overhead pairs patent peters position positioned potential potentially prior problem processing prospective prospectively protocols pulse pulses radians radius rads raster rectangular represents required resolution retrospective retrospectively rewinds sample samples schedule sensitivity sick slab sometimes space spatial spin subset sufficient table tables takes term theory train trains trajectory treated types variable varied versa vice waveforms words zero