Abstract #2561

A Fast and Robust Method for Off-Resonance Detection in Metal Implant Imaging

Theresa Bachschmidt1, 2, Mathias Nittka1, Peter M. Jakob2

1Siemens Healthcare Sector, Erlangen, Germany; 2EP5, University of Wuerzburg, Wuerzburg, Germany

Field inhomogeneities are the dominant source for artifacts in images of volumes containing metal implants. They can be determined by measurement of the geometric distortion of the slice profile. Its shape is determined by the amplitude and polarity of the readout gradient. This work depicts analytically two inaccuracies of this method, both based on the fact that slices are not infinitely thin and suggests solutions. Application of readout gradients with amplitudes in the range of the slice select gradient require opposite polarity relative to the slice select gradient. The application of two different readout gradients can resolve the inaccuracy of the excitation bandwidth.

Keywords

ability according accuracy achieved acquisition adapting additional advantageous aliasing allows although ambiguity amplitude amplitudes application applied applying artifacts automated avoid axis bandwidth bottle bottom calibration chromium clear cobalt combination compromised consuming containing contribute cover decisive denotes depict desired detect detected detection determined determines disambiguation distance distortion doped eliminates employed encoding ensured equal equation except excitation excited expressed extended fast feasible femoral field frequency full function future general geometric gradient gradients hence identified implant implants implemented improved indicate individual individually inhomogeneities introduce inversion inverted keep latter measured metal modified noise note occurs offset operative opposite part parts perturbations peter phantom place plastic polarities polarity possibility post predicted prior process profile profiles projection prolonged proposed quantification readout reduce reduced refinement related removes required restricted risking robust sampling scanner scout sector select selected sensitive separate severely since slice spatial specified spectrum steps structures subject suggested tailored titanium true turbo upper vicinity visualization water words