Abstract #2560

Assessing Measurement Accuracy Near Orthopaedic Materials in Magnetic Resonance Imaging

Matthew F. Koff1, Parina Shah1, Kevin M. Koch2, Hollis G. Potter1

1Department of Radiology and Imaging, Hospital for Special Surgery, New York, NY, United States; 2Applied Science Laboratory, General Electric Healthcare, Waukesha, WI, United States

The presence of orthopaedic hardware creates significant distortion artifacts in MR images. This study evaluated the accuracy of artifact reduction between standard of care 2D fast-spin-echo (FSE) images and corresponding MAVRIC images as compared to a gold standard. Metal samples produced significant distortions in 2D-FSE images, which were eliminated in MAVRIC images. The average difference between the known phantom data points and MAVRIC data points was less than 2 pixels for scans of all materials. MAVRIC scans are effective in reducing image distortion, when compared to known dimensions.

Keywords

accuracy acquisition acquisitions adverse aided aligned allows amount applied approximate array artifact audience automated available bandwidth beneficial calculation care cause channel chrome clinical clinically cobalt coil combination combining common commonly components composed confirm control coordinate coordinates corner corners coronal dataset datasets debris design determine deviations dimensions direct distance distances distortion effectively electric encoding evaluated exact excitation excluded existed fast findings frequency function future general generated geometric geometry gold grid ground header identical identified implant importance indicate individual inside institutional joint known laboratory limited local location locations material materials matrices matrix minimizes model molecular near novel numerous obscuring pathology phantom pixel pixels polyethylene potter precision prevented previously program protocol radiology reactions reduced reduction refocusing repeated replacements representative response sample samples scanner science semi series shah since slice soft special specifically stainless steel studies subdivided subregions superior support surgery system target tissue tissues titanium truth understanding uniform utilized variable visible visualization volume wear