Abstract #1589

Clinical Evaluation of High Permittivity Pads for Improving Abdominal Image Quality at 3T

Gunhild Erstad Aandal1, Gurpreet S. Sandhu1, Ronald D. Novak1, Dean Nakamoto1, Luis Landeras1, Mark A. Griswold1, 2, Paul de Heer3, Andrew G. Webb3, Vikas Gulani1, 4

1Dept. of Radiology, University Hospitals Case Medical Center, Cleveland, OH, United States; 2Dept. of Biomedical Engineering, Case Western Reserve University, Cleveland, OH, United States; 3Dept. of Radiology, Leiden University Medical Center, Leiden, ZA, Netherlands; 4Dept. of Biomedical Engineering , Case Western Reserve University, Cleveland, OH, United States

3T abdominal imaging is limited by image quality loss due to inhomogeneities in the transmitted RF field, even with dual channel transmit. Recent simulation and in vivo volunteer studies have shown that thin pads made with novel high permittivity materials are a simple, inexpensive solution to ameliorate these problems. Here we show pilot patient studies comparing image quality without and with these pads, for 3D fat suppressed gradient echo imaging of the abdomen in a dual transmit 3T system. The preliminary data confirm the potential of these pads for use in improving image quality for abdominal imaging at 3T.

Keywords

abdomen abdominal acquisition actively address affecting already alternative amelioration another anterior archive around assess assigned audience background barium better biomedical board body centrally channel chloride choice clinical clinicians communication compliant conducted considered critically cumbersome dean definitive degree dept described describing developed deviation diagnostic dielectric differential dimensions dissolved dual eliminated engineering evaluation expensive explore explored extremely field forced fundamental gain geometries hence homogeneity implication important improve improved improvement improving inhomogeneities inhomogeneity initial investigated issues larger lesion local made males mark measured median medical migration mild mildly mixed moderate modifications needed pads paired particularly patients perceived picture pilot pixel place population potential preference preferred previously problem profound quality quantified radiologists radiology readers reason recently reducing remaining respectively reviewer reviewers scale score scores severe side significantly simply simulation spatial specialized station studies subject subsequently sufficiently superior suppressed system systems target thick thousands titanate towards transmit trend uniformity utilized varying vibe vicinity wall western widespread