Abstract #1396

Improved Adiabatic Inversion Design for Myocardial T1-Mapping

Peter Kellman1, Daniel A. Herzka2, Michael Schacht Hansen3

1NHLBI, NIH, Bethesda, MD, United States; 2Biomedical Engineering, Johns Hopkins University, Baltimore, MD, United States; 3NHLBI, National Institutes of Health, Bethesda, MD, United States

Myocardial T1-mapping based on Look-Locker methods such as MOLLI rely on ideal inversion recovery. Adiabatic inversion pulses used to mitigate inhomogeneity of transmit field strength do not achieve perfect inversion as a result of transverse relaxation (T2) during the pulse. Imperfect inversion leads directly to a T2-dependent error in the estimate of T1. An improved adiabatic inversion pulse optimized for myocardial T1-mapping was designed and evaluated experimentally. Due to peak power constraints, a tan/tanh design was found to achieve the best inversion performance for a given design specification. Reduced dependence on T1 and T2 facilitate a calibrated correction of T1-estimates.

Keywords

achieve achieves adiabatic agar amplitude approx assumes bandwidth beta better biomedical brute calibrated cardiovascular complete concentrations condition considered constrained constraints contours correct correction dashed denoted dependence dependent design designed designs detecting directly doped dotted dual duration edema elevated engineering ensure equations error evaluated existing expected experimental experimentally exponential facilitate fibrosis field force frequency graphed greatly green ideal illustrating imperfect improved included independence indicated inhomogeneity inversion johns leads locker look mapping measured medical medicine mitigate myocardial myocardium note optimization optimized peak perfect performance peter phantoms plot power product pulse pulses quantifying recovery reduce reduced reducing rely response school several shape shorter solid solutions space strength sweep transmit transverse uncorrected validation variation various varying zeta