Comparison of Myocardial T1-Mapping Protocols: Accuracy and Precision

Peter Kellman1, Hui Xue2, Michael Schacht Hansen3

1NHLBI, NIH, Bethesda, MD, United States; 2Siemens Corporate Research, Princeton, NJ, United States; 3NHLBI, National Institutes of Health, Bethesda, MD, United States

Both accuracy and precision of T1-mapping are important for quantitative measurements and reliable detection of abnormal elevation of T1. Accuracy reflects the systematic or bias errors while precision reflects the random error due to noise. The accuracy and precision of several popular methods, including inversion recovery and saturation recovery schemes, are investigated in detail using a waveform level Bloch simulation to assess accuracy, and Monte-Carlo method of repeated trials to assess precision. Simulations are confirmed with phantom measurements.

Keywords

abnormal accuracy achieved achieves acquiring acquisition adjustments advantage affected agar alternative apparent applies approaches assess assumed best bias bottom called cause chow compute concentrations confirm confirmed conjunction considerably consistent constant contrast corporate correction cost curve decreased degrees depend detail detect detection determine deviation deviations diffuse disease doped dynamic elevation eliminated eliminating equations error errors expected experimental fitting focal freedom frequency function good greater heartbeat heartbeats ideal illustration important improved in vivo include incorporating influence initial inversion investigated known largely larger locker look loss magnetization mapping maps medical minimizing model myocardial myocardium native need noise offs optimized originally peter phantom phantoms physics pixel plus popular post precision predicted preparation prepped processes proposed protocol protocols quantitative random readout recovery reduced reducing referred reflects relatively repeated reported respectively sampling saturation scanner scheme several shorter shot simulation simulations slightly standpoint steady subsequent subsequently systematic theory timing tissue trade trials tube upper validation various varying waveform whereas worse