Meeting Banner
Abstract #0767

Decoupling of Tx/Rx Coils Using a Tx-Array System: Application to UTE and CEA

Ali Caglar Ozen1, 2, Ergin Atalar1, 2

1Electrical and Electronics Engineering, Bilkent University, Ankara, Turkey; 2National Magnetic Resonance Research Center (UMRAM), Ankara, Turkey

A novel method for decoupling that provides over 80dB isolation between receive and transmit coils is developed. Transmit array technology is utilized to cancel the individual B1 induced currents in the receive coil by adjusting relative amplitudes and phases of transmit array elements. As an application of this method, MRI with concurrent excitation and acquisition (CEA), where MR signal is acquired during RF excitation, is implemented and spectroscopy and frequency-swept imaging techniques are demonstrated. CEA enable observation of spin characteristics during RF excitation as well as zero-TE. Preliminary results are represented for spectroscopy and imaging of ultra-short T2 samples.

Keywords

achieved acquisition advantageous advantages afterwards alcohol amount amplitude amplitudes applications applied approaches array artifact artifacts bandwidth becoming bonding built channel channels characteristics clinical coil coils commonly complex component considered continuous coupler deconvolution decoupling decreased delay dependent detuning developed diameter diodes distinctive distortions done driven duration dynamic edges electrical electronics elements eliminates employed epoxy ethyl except excitation expected experiment experimental expressed fails fast field frequency grid helps holes home implementation include increasing individual induced induces inside isolation know leakage ligaments loss magnitude mapping measured missed modes modulator need noise noted novel observe observing onto output paper peak phantom placement plate popular position powers problem procedure projections providing pulse pulsed radial radio rapid reason receive receiver reconstruction recorded reducing remaining require rubber samples short since solid solution solve sophisticated space specified spectroscopy spectrum spin spins starts still structures subtraction surface sweep system systems tendons trajectory transmit true turkey ultra uniformly voltage wave zero