Fast Imaging of Metallic Implants by Data Subsampling

Bragi Sveinsson1, Pauline W. Worters2, Garry E. Gold1, Brian Andrew Hargreaves1

1Stanford University, Stanford, CA, United States; 2GE Healthcare, Menlo Park, CA, United States

Slice Encoding for Metal Artifact Correction (SEMAC) corrects for magnetic field variations by applying view-angle tilting in the readout direction to avoid in-plane distortion, and phase encoding in the slice direction to resolve distorted slice profiles. Phase encoding in the y and z directions can lead to long scan times. We demonstrate a way to reduce the scan time by 50% by subsampling ky-kz space in a checkerboard fashion. The resulting aliased copies do not interfere with the desired image because of the shape of the slice profile and can be removed during post-processing. We demonstrate the feasibility of this approach on a phantom and for in vivo scans, in combination with partial Fourier and parallel imaging.

Keywords

ability acceleration achieved agar aliased aliasing allow allows amount applied applying around arrow arrows artifacts assuming assumptions bone calibration cause central checkerboard close closely combined comparable compatibility complications contains conventionally copies correct corrects currently dashed date demonstration desirable desired detecting detection distorted distortion distortions done effective efficient encoded encoding energy equally even excited extent fast feasibility field finally fixation full fully funding gold gradient hardware highly immersed implant implants improve improvements in vivo include interfere joint keeping largest lead limited long makes manner many metallic minimal modified near next otherwise outside overall packed packing packs parallel park partial particularly patients pattern permits portion post prevalence processing profile prospectively prosthesis quality recombined reconstruction rectangular reduced reduces reduction removed replicas requires resolving rest retroactively retrospectively sampled sampling saving scanned scanning scheme severe shoulder since slice space sparse spectral spent spinal strengths subject support system theory tilting titanium twice twofold typical unused view volume volumes zeroed