T2-Weighted BOLD fMRI at 9.4 T Using a S2-SSFP-EPI Sequence

Philipp Ehses1, 2, Juliane Budde1, Gunamony Shajan1, Klaus Scheffler1, 2

1High-Field MR Center, Max Planck Institute for Biological Cybernetics, Tbingen, Germany; 2Dept. of Biomedical Magnetic Resonance, University of Tbingen, Tbingen, Germany

It has been hypothesized that T2-weighted BOLD fMRI at ultra-high field shows higher spatial specificity than T2*-weighted BOLD, since the main signal contribution is expected to come from the extravascular spins of the microvasculature. Unfortunately, the number of slices that can be acquired in multi-slice SE-EPI is limited at ultra-high field due to SAR constraints. T2-weighted steady-state sequences, such as S2-SSFP, have been previously used as an alternative to spin-echo based BOLD fMRI. In this work, we present a 3D S2-SSFP sequence that is accelerated using an EPI readout and show first results from finger tapping experiments at 9.4 T.

Keywords

accelerated acquisition activation activations additional advantage allow allowed alternative analog anatomical appear applied approval approx artifacts balanced bingen biological biomedical blips bold brain built caused central channels chemical coil come committee comparable comparing consent consisting constraints contribution correction cortex custom cybernetics datasets dept developed distortion distortions encoding ethics examine excitation expected extraction feat field filtering finger fold frames function functional hand head healthy highest highly hypothesized imperfect indicates inflow informed institute isotropic latter limitations limited local long maps miller monitor motor necessary often outer overlay oversampling paradigm partial preliminary prevent previously problem processed profile proposed readout receive reception registered repeated representative required residual resolution response rest seems sensitivity short shorter simple since slice slices slightly smoothing spacing spatial specificity spin spins spread steady step steps strong stronger tapping temporal train transmission transmit twice ultra unbalanced unfortunately vein veins viable volume volumes volunteer whether