Improvement in B1+ Homogeneity of 3T Cardiac MRI in Swine with Dual-Source Parallel RF Excitation

Daniel A. Herzka1, Haiyan Ding2, 3, Michael Schar4, 5

1Biomedical Engineering, Johns Hopkins University, Baltimore, MD, United States; 2Biomedical Engineering, Tsinghua University, Beijing, China; 3Biomedical Engineering, Johns Hopkins School of Medicine, Baltimore, MD, United States; 4Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins School of Medicine, Baltimore, MD, United States; 5Clinical Science, Philips Healthcare, Cleveland, OH, United States

MRI scanners use an integrated birdcage coil to generate radio frequency (RF) excitation fields (B1+). It has been reported that at 3T variations in flip angle can range from 31 to 66% (B1+ inhomogeneities) over the left ventricle and average flip angles can be reduced by ~20% (RF power). Multi-channel transmit systems allow RF-shimming to locally improve the B1+ field. We quantify improvements in B1+ field homogeneity and average RF power resulting from dual-source parallel RF excitation in cardiac swine imaging. Correlation with animal size demonstrates that larger subjects suffer more B1+ field inhomogeneity and benefit more from RF shimming.

Keywords

achieved affected aimed analyzed animal animals array artifacts bars best biased biomedical blue body bottom breakdown breath cardiac channel characterized china clinical coefficients coil consistent correlated correlation cover cross days depend desired dielectric distribution drawn dual effective ellipse elliptical encompassing engineering entire erroneous error estimation eventually example excitation expected field fields frequency guide heart heterogeneity homogeneity homogeneous humans implies improvement improvements indicate indicating inhomogeneities inhomogeneous involves johns larger lateral lead local localize localized loss magnetization magnitude manually mapping maps measured measures medicine months overall paired parallel part penetration percent percentage performance pixel pixels placed position positive positively post power predictability prep preparation pulses quantify quantitative radiological radiology ranging reduce reduces reduction repeatedly represent requested respectively resting sample saturated scanners school science sectional sessions shading shimming significance significantly since slice source sources strength studies sung superimposed swine system systems tailed though transmit triggered twenty typical variation variations ventricle ventricular volume whole whose