Improved Excitation Fidelity in Cardiac Imaging with 2-Spoke Parallel Excitation at 7 Tesla

Sebastian Schmitter1, Lance DelaBarre1, Xiaoping Wu1, Andreas Greiser2, Dingxin Wang3, Kamil Ugurbil1, Pierre-Francois Van de Moortele1

1Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, MN, United States; 2Siemens AG, Erlangen, Bavaria, Germany; 3Siemens Medical Solutions USA, Inc., Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, MN, United States

Cardiac MRI may greatly benefit from ultra-high field (UHF) providing higher SNR and intrinsic contrast. However, shorter RF wavelength at UHF results in transmit B1 (B1+) and contrast variations through the heart. This problem has been addressed using multi-channel transmit coils and B1-shimming, but further improvements are expected by using parallel transmission (pTX) with multi-spoke RF-pulses as shown in brain and liver at 7T. However, this involves additional challenges, including rapid and robust multi-channel ECG-triggered B1+ calibration, and sensitivity to motion. Here, we investigate the impact of 2-spoke RF-pulse design on 7T cardiac imaging using a 16-channel pTX system.

Keywords

absence absolute acquisition acquisitions additional agreement aiming although amplitude aorta applied applying approved axial become benefit body bottom brain calibration cardiac cardiovascular challenges challenging chamber channel channels cine coil columns combination computed conservative consistent consistently contrast corrected correspond cycle delay design designed desirable diastole difficult drawn dropping duration edges enabling energy equal especially estimation even excitation excitations existing expected fairly fast feasible fidelity field final fixed foundation full good gradient gray healthy heart human imposes improved improvements improves included inhomogeneities inhomogeneity initial investigate investigations involved just largely last limiting magnet magnitude manually maps measured mode modified motion multichannel needed note optimization optimized organ organs overall parallel pattern patterns peak posterior posteriorly predictions preserving profile providing pulse pulses ramps rapid rapidly reduced residual robust safety scale sensitivity shape shimming shorter significantly signing slight solutions spoke spokes successfully system target technical though transmission transmit trigger triggered ultra variation variations view views volunteers wavelength