Meeting Banner
Abstract #4040

Fat- And Susceptibility-Corrected R2* Mapping for Liver Iron Quantification: Preliminary Evaluation in a Healthy Cohort

Diego Hernando1, Jens-Peter Khn1, 2, Scott B. Reeder1, 3

1Department of Radiology, University of Wisconsin-Madison, Madison, WI, United States; 2Department of Diagnostic Radiology, University Greifswald, Greifswald, MV, Germany; 3Department of Medical Physics, University of Wisconsin-Madison, Madison, WI, United States

R2*-MRI can be used to assess body iron deposition for diagnosis, staging and treatment monitoring of iron overload. However, the apparent R2* is often confounded by the presence of fat, noise floor and susceptibility effects (e.g., near tissue/air interfaces). Recently, a method was introduced to perform fat-corrected, complex-fitting, and susceptibility-corrected R2* mapping. In this work, we quantitatively evaluate the susceptibility correction in a cohort of 35 healthy subjects without iron overload. Uncorrected measurements showed significant deviations in R2*. These deviations were largely corrected by susceptibility-corrected R2*. This technique may be important for the assessment of iron deposition in the liver.

Keywords

accelerator accounting acquisition acquisitions affect affected agree agreement although anatomy apparent appear approval artifacts assess assessed avoid axial blood blue body cohort complete complex components confounding contributing corrected correcting correction correlates demonstrating dependent deposition described diagnosis diagnostic dimension discarded diseased distorted dome elevated errors estimation evaluate evaluation example exams except field filter findings fitting floor foundation good gradient gradients healthy highly homogeneous important improvements improves included increasing individual introduced invest iron lead liver livers lobe localized longer mapping maps measured measuring minor modeling monitoring motion near noise often optimized overload partial particularly pathologic patient peak physics plus precision precluded presence primarily problematic produces proposed quantification radiologist radiology rapid rapidly recently reconstructed reconstruction reconstructions reeder remove repeatability representative resection respectively scanned segment segments several severe slice slices slowly space spatial spoiled staging subject subjects support susceptibility threshold trains transplant treatment twice uncorrected upon variation variations varies volumetric volunteer whole windowing wood years