Knee Cartilage T1 Mapping with High Resolution Multi Slice Inversion Recovery: Feasibility, Reproducibility and Accuracy

Henk Smit1, J. van Tiel1, E.E. Bron1, D. Poot1, G.C. Houston2, W.J. Niessen1, 3, H. Weinans1, G.P. Krestin1, S. Klein1, E.H.G. Oei1, G.

1Erasmus MC, Rotterdam, Zuid Holland, Netherlands; 2GE Healthcare, Netherlands; 3Delft University of Technology, Netherlands

In conventional T1 mapping of the human knee cartilage the low resolution is a problem. We propose and evaluate an alternative T1 mapping technique of the cartilage. Multi-Slice Inversion Recovery with ARC parallel readout allows high resolution knee cartilage images in a short time. Also a method for error estimation of the derived T1 values is presented, which is used to compute a weighted mean T1 over a region of interest. Reproducibility is tested on healthy volunteers and sensitivity to cartilage damage is shown on an osteoarthritis patient.

Keywords

absolute accounting accuracy acquisition addition allowing alternative although applied arrows artifacts bearing bounds calibration cartilage cause clinical clinically commonly compute computing contact correction damage dead delayed delft derived dimensional distribution double doubling drawn effective enhanced error estimation estimator evaluate evaluation fast feasibility feasible femoral femur furthermore gadolinium gradient healthy highly hires human improved in vivo increasing indicating inversion journal knee lateral layer likelihood listed local localized locker longest look magnitude mapping maps mask masks materials measure medial metric moreover motion mutual noise osteoarthritis outermost overlay parallel partial parts pathology patient pattern plateau possibility posterior produced propose protocol protocols quantification readout recovery reduce reduces registered repeatability reproducibility reproducible resolution rigidly root scanned scanner selected short shortened slice spin spoiled square suffer suitable table technology thin tibial topography twice typical typically uncertainty variable versus visible visualized vitro volume volunteer volunteers zooming