Mehmet Arcan Erturk1, 2, AbdEl-Monem M. El-Sharkawy2, Paul A. Bottomley, 12

1Electrical & Computer Engineering Department, Johns Hopkins University, Baltimore, MD, United States; 2Department of Radiology, Johns Hopkins University, Baltimore, MD, United States

Recent work with interventional loopless antennae suggest a near-quadratic gain in signal-to-noise ratio (SNR) with field strength, and potentially safe operation at higher fields (B0>1.5T). This work reports the theoretical and experimental SNR, and initial safety testing of the first intravascular loopless antenna at 7T. An absolute SNR gain of 5.71.5-fold was realized at 7T vs. 3T: more than 20-fold higher than at 1.5T. The probe heated <2.0˚C during 15min RF exposure at an applied 4W/kg SAR. The SNR gain offers the potential for intra-vascular MRI microscopy at <<100μm resolution, larger fields-of-view (FOV) and/or high-speed MRI endoscopy.

Keywords

absent absolute absorption agree anatomy antenna antennae applied arterial artery autopsy axial bath biological black blue cables calcification carotid circuitry circular coil colored comparable computations computed computer conductivity conductors confirmed connected constant continuous contours corrected cyan dark decoupling detector determine determined development device diameter dielectric driven effective electrical electromagnetic endoscopy engineering exceed excitation experimental experimentally exposure extended extra fabricated field finally fixed fold form gain gains geometries geometry green head highest human identical inner inside institution intact interface internal johns junction least lengths local locations long measured microscopy minutes moments noise normalization normalize numerical numerically offering offers operated parallel performance phantom phantoms phys placed plaque potential potentially preamplifier press prior properties pulses purple quadratic radiology recorded report resolution resonant room safe safely safety saline scanner screen sensors service sites south specimen specimens speed spin suggest suggests supported temperature theory thermal tissue tune turbo uniform usage useful view vitro wave whips whose