New Insights Into PET Count Rate Reduction During Simultaneous MR-PET Measurements

Daniel Brenner1, Christoph Weirich1, Jrgen Scheins1, Etienne Besancon1, Lutz Tellmann1, Hans Herzog1, N. Jon Shah1, 2

1Institute of Neuroscience and Medicine - 4, Forschungszentrum Jlich, Jlich, Germany; 2JARA - Faculty of Medicine, RWTH Aachen University, Aachen, Germany

Minor influences of MRI on the PET detectors during simultaneous MR-PET measurements with the 3T MR-BrainPET have been observed and shown to result in a drop of the measured count rate of up to three percent. This study analyses the distortions generated in the output signal of the PET detector by the time varying MR gradients. Pronounced effects are visible at start and beginning of the gradient ramps while an increased noise level was observed during the gradient flat top.

Keywords

active additionally aforementioned allow amplification amplitude analogue assumed becomes block caused causing characteristic considered count created custom cyan cycles dedicated definition depend describe design detailed details detector detectors developed distorted distortion distortions drop duration eddy electronics events evident examinations execution faculty fields flat frequency gradient gradients green hypothesis indicate induced influence influences insight insights institute largest linearly macroscopic magenta measured medicine might minor neither noise noteworthy observe origin oscilloscope peaks percent process processing pulse pulses ramps received reduce reduction repetition report responsible revealed rise scintillation separate served shah shaping shorter simultaneous smoothed spectrum spikes start strengthens studies switching targeting trigger varying visible waveform yellow