Building a Combined Cyclotron and MRI Facility: Implications for Interference

Mark BM Hofman1, Joost PA Kuijer1, Jan Willem de Ridder1, Lars R. Perk2, Rudolf M. Verdaasdonk1

1Physics and Medical Technology, VU University Medical Center, Amsterdam, Netherlands; 2BV Cyclotron VU, Amsterdam, Netherlands

With the introduction of hybrid PET/MRI systems, it has become more likely that cyclotron and MRI systems will be located close to each other. This study considered the theoretical interference of a cyclotron facility on a MRI system. Measurements and simulation showed that the magnetic field of a cyclotron can be considered a magnetic dipole field. In a specific case of an 18 MeV cyclotron and a 3T superconducting whole body MR system a minimum distance of 11-21 m (depending on location and criteria applied) has to be considered to prevent interference.

Keywords

acceptable account active agree allowable allowed applied apply approximated around assessed assumed become besides body building built cage characterized clinical close closely closer coil combined companies compensated compensation component components consider considered contain creating criterion critical cyclone cyclotron cyclotrons depending derivatives derived described determined diameter dipole direct directly discovery distance distances drop dynamic dynamical electromagnet equations especially even exclude external facilities facility field fields frequency general gradient gradients guidelines hours index induction installation interference larger located location locations loop magnet magnetometers magnets mainly manual manufacturer manufacturers mark measure measured medical micro moment much mutual newly offer office outside parallel passive past peak perfectly perturbation physics planned planning process ramping realistic setup several shielded shielding shorter simulated simulating simulation site situ situation sources spatial specifications specify spectroscopy steeper stray strength strengths strong superconducting survey switched switching system systems take target theoretical theory type typical typically versus vertical view whereas whole width zero