PROPELLER with Echo Stabilization

ZHIQIANG LI1, Dawei Gui1, Ajeet Gaddipati1, Xiaoli Zhao1

1MR Engineering, GE Healthcare, Waukesha, WI, United States

PROPELLER has been widely used in the clinic, but also experiences the violation of the Carr-Purcell-Meiboom-Gill condition as in FSE, which degrades the signal. FSE employs phase correction to measure the phase error and adjust the data acquisition accordingly. However, its direct application in PROPELLER is not feasible since measurement at all blade angles leads to unacceptable increase in scan time. In this work, a model based on physical gradient is proposed so that the phase error can be predicted from measurement at a few blade angles. This reduces scan time, helps the application of PROPELLER in body/MSK imaging.

Keywords

abstract acceptable accurately acquisition actual amplitude anatomies application appropriate artifacts artificially asterisk asterisks axis behavior better blade blades blue body capability caused causes choosing clinic clinical coefficients compensate compensates condition constant consuming contributed correcting correction dependence depict desirable eddy employed engineering environment error errors especially estimation experimental extended faces fidelity field generalized generate gill gradient greatly improve improves in vivo inherent inhomogeneity insensitivity issues kept limited match measure measured model motion needed nonlinearity noting partially performance physical pipe polarity predict predicted propeller propose proposed quality recently respect rotating rotation satisfy self sets since sinusoidal solution source stabilization stabilize straightforward strictly studies subject summary susceptible theory throughout train utilizes various violation widely worth wrist