Abstract #2216

B1+ Inhomogeneity Compensation for RF Refocusing Pulses in Spatially Selective Excitation (SSE) at 7T

Tomasz Dawid Lindel1, 2, Andre Kuehne1, 2, Patrick Waxmann1, 2, Frank Seifert1, 2, Thoralf Niendorf2, Bernd Ittermann1, 2

1Medical Metrology, Physikalisch-Technische Bundesanstalt (PTB), Abbestr. 2-12, 10587 Berlin, Germany; 2Max Delbrck Center for Molecular Medicine (MDC), Berlin Ultra-High Field Facility (BUFF), Berlin, Germany

For brain imaging two-dimensional spatially selective excitation using parallel transmission (pTx) is a well-established technique these days. The missing selectivity along the third spatial dimension is a severe limitation, however. To overcome this we combined 2D-SSE with a conventional slice selective refocusing pulse for phantom and in vivo imaging at 7T. Initially, this resulted in severe signal inhomogeneities due to the spatially non-uniform refocusing at the ultrahigh field. Using RF shimmed refocusing pulses combined with adapted pTx pulses the problem could be resolved.

Keywords

acceleration achieve adaptation adapted allows amplitude anticipating apex array artifacts avoid better biomedical body buff calculation channel clean clearly clinical coil combined common compensate compensating compensation conditions constant corrective correspondingly cross cylindrical define definition described determined detrimental dimension disadvantage distorting distortions easily efficacy eight element ending ensure equipped especially established excellent excitation excite existing expensive experiment facility feasibility field fields flat fourfold frank good hardware head healthy homogeneity homogeneous homogenization human imperfections implemented improves in vivo induced inhomogeneities inhomogeneity issues just length limited maps measured measures medical medicine mode modulation molecular much nominally often panels parallel pattern performance phantom phantoms planar powerful prior promising pulse pulses quadratic quality rapid refocusing removes residual restrict scanner scanners scheme sections selection selective selectivity sense severe shimmed shimming significantly slew slice space spatial spatially spiral square starting still subsequent sufficient target third transmission turn ultra ultrahigh unsatisfactory variation visible vitro volunteer whole yellow