Abstract #1995

Parallel Transmit Vessel Selective Arterial Spin Labelling: Phantom and In-Vivo Results

MAGNA25Aaron Oliver-Taylor1, Chris Randell2, Roger J. Ordidge3, David L. Thomas4

1Department of Medical Physics and Bioengineering, University College London, London, England, United Kingdom; 2PulseTeq Products Division, Renishaw PLC, United Kingdom; 3Centre for Neuroscience, University of Melbourne, Melbourne, Victoria, Australia; 4Institute of Neurology, University College London, London, England, United Kingdom

Vessel selective continuous arterial spin labelling can be achieved using a single surface coil placed over one of the carotid arteries. However the labelling B1 field extends across the neck causing contralateral labelling. Presented is a method using parallel transmission to nullify the B1 field at the contralateral artery. Both phantom and in-vivo results show a reduction in contralateral labelling is possible, improving specificity without losing the advantageous high labelling efficiency and no magnetisation transfer effects of separate-coil CASL.

Keywords

accelerated additional amplitude amplitudes appropriate array arterial arteries artery body built carotid central channel circular circulated coil coils college comprising computer contaminated continuous correction created cycled decrease delay dell deposition depth diode distribution division doped driven drove dual duration efficiency either element england ensure equipped exactly extends feeding female field filter fine flow gradient gradients head healthy hemisphere histogram house improve improving in vivo include independently individual inflow interface interfaced inversion kingdom label linear locate manual manually maps masks mass matched matrix minimal much neck neurology nominal null oils oliver opposite parallel partial perfusion peristaltic perturbation phantom physics pipe pipes placed positioned post power produce products profile pulses quantification rand reception reduce reducing reduction required retain retaining roger segmented select selective selectively sett side simulations sing slabs slice spatial specificity spin suppression surface surrey system targeted territory transfer transformed transmit transverse trio tuned user vertebral vessel volumes water whilst whole year yield zero