Meeting Banner
Abstract #1565

MR Safety: Fast T1 Thermometry of the RF-Induced Heating of Medical Devices

Daniel Gensler1, Florian Fidler1, Philipp Ehses1, Marcus Warmuth2, Theresa Reiter2, Markus Dring3, 4, Oliver Ritter2, Mark E. Ladd5, Harald H. Quick6, Peter M. Jakob1, 3, Wolfgang R. Bauer2, Peter Nordbeck2

1Research Center Magnetic-Resonance-Bavaria, Wrzburg, Germany; 2Department of Internal Medicine I - Cardiology, University Hospital Wrzburg, Wrzburg, Germany; 3Experimental Physics 5, University Wrzburg, Wrzburg, Germany; 4Noras MRI Products GmbH, Hchberg, Germany; 5Erwin L. Hahn Institute for Magnetic Resonance Imaging, University Duisburg-Essen, Essen, Germany; 6Institute for Medical Physics, University Erlangen-Nrnberg, Erlangen, Germany

Determining the MR compatibility of medical implants is becoming increasingly relevant. In most cases, the RF-heating of implants is measured by fluoroptic probes, but these can only measure the temperature at a single point. Another method to determine heating effects is MR-thermometry using the PRFS which gives good results in homogeneous phantoms. However, in several cases the inhomogeneity of organic tissue and susceptibility changes near an implant prohibits PRFS-thermometry. The intention of this work was to develop a fast T1-based thermometry method which allows controlled MR-related heating of a medical implant while simultaneously quantifying the spatial and temporal temperature distribution.

Keywords

absorption acrylic aging allowed allows another ants applicability applicable applied artifacts baron black blue body cable capable cardiology caused cellulose chosen coil comm compatibility computed conductive controlled course covered cycle defined delay determine develop deviation devices dots drawback dynamic early eating even every examined example excellent experiment fast field fitted flash general generate germ gives good guided hand heating hence hods homogeneous implant implants implicit important in vivo included induced initial institute intermediate inversion magma major many maps mate measured medical meet meters ming minutes muscle near needle next often oliver organic overlay part particularly peter phantoms physics piece pixel plants pork position poster power powerful presence probe products promising proof proposed proscribe proton pulse pulses quantification quantify quantifying quick quired related relevant resolution ring schematic sensor several situations slice sliding spatial spatially studies subtracting surface susceptibility system temp temperature temporal therm thermometry tissue titanium tool validated versus vices view vitro whole window