Water Specific Magnetization Transfer in Skeletal Muscle Using MT-IDEAL

Robert L. Janiczek1, 2, Christopher D.J. Sinclair2, Giulio Gambarota3, Xavier Golay2, Rexford D. Newbould4, John S. Thornton2

1Global Imaging Unit, GlaxoSmithKline, London, United Kingdom; 2Institute of Neurology, University College London, London, United Kingdom; 3Pharmaceutical and Biological Sciences, Universit de Rennes 1, Rennes, France; 4Imanova Ltd., London, United Kingdom

A reduced magnetization transfer (MT) effect has been observed in muscles affected by neuromuscular diseases due to a combination of lipid infiltration, edema, and/or macromolecular differences relative to healthy muscle. Lipid infiltration can therefore mask MT changes due edema and biologically important macromolecular abnormalities. This work presents a technique, MT-IDEAL, that combines a chemical-species separation imaging technique, IDEAL, with an MT imaging acquisition. MT-IDEAL uses multiple echoes to increase SNR as well as produce fat fraction, T2*, and water-isolated MTR maps with no scan time penalty.

Keywords

abnormalities accounts achievable affected afforded alongside arrow asymmetry body chem collected college combination complex concentration content deal disease echoes edema enabling equation equations experiment fatty fraction generated gradient gray greater healthy idea ideal inclusion infiltration institute isolated iterative john kingdom least lipid lipids local longer mapping markedly masks measured might mined minim modeled muscle neurology neuromuscular panels partial patient peaks pharmaceutical pixels previously pulse ranging recovery reduced reeder remove respectively sciences source species term trio unique unit water white