Abstract #1182

4D Phase Contrast MRI in Intracranial Aneurysms: A Comparison with Patient-Specific Computational Fluid Dynamics with Temporal and Spatial Inflow Velocity Boundary Conditions as Measured with 2D Phase Contrast MRI

Pim van Ooij1, 2, Joppe Schneiders1, Henk A. Marquering1, 2, Charles B. Majoie1, Ed vanBavel2, Aart J. Nederveen1

1Radiology, Academic Medical Center, Amsterdam, Netherlands; 2Biomedical Engineering & Physics, Academic Medical Center, Amsterdam, Netherlands

4D phase contrast MRI measurements (4D PC-MRI) in intracranial aneurysms are presented and compared with patient-specific computational fluid dynamics in which a 4D velocity profile as measured with through-plane PC-MRI (2D PC-MRI) is prescribed as inflow boundary conditions. Quantitative and qualitative similarities were found. Although the inflow velocity assumption in CFD is minimized by using the 2D PC-MRI measurement, other assumptions cause deviations from measured flow patterns. However, in CFD the resolution is higher and results show more flow details. Thus, for hemodynamic assessment, CFD and 4D PC-MRI complement each other.

Keywords

anatomy aneurysm aneurysms angiography approved artery assumption assumptions attempting attributed best biomedical booth boundary cardiac coil column conditions consent contra contrast contribute corresponded days details deviate dies engineering ethics evolution faces features field flow fluent fluid formation geometric geometry head impingement inflow insufficient last local lumen magnitude measured median medical meshes minimize models much natl noise onto paired patient patterns perpendicularly physics placed predict preferred prescribed profile radiology reflected registered resolution resolved retrospective risk rupture semi sense sett shear similarities simulation simulations slice spatial step stress stud subj table temp temporal unsolved vectors velocities velocity viscosity wall ween whereas wise