Meeting Banner
Abstract #0694

Improved Q-Ball Imaging Using a 300 MT/m Human Gradient

SUMMA25Julien Cohen-Adad1, M Dylan Tisdall1, Ralph Kimmlingen2, Eva Eberlein2, Thomas Witzel1, Philipp Hoecht2, Boris Keil1, Juergen Nistler2, Dietmar Lehne2, Keith Heberlein3, Jennifer A. McNab1, Herbert Thein2, Franz Schmitt2, Bruce R. Rosen1, Van J. Wedeen1, Lawrence L. Wald1, 4

1A. A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Harvard Medical School, Charlestown, MA, United States; 2Siemens Healthcare, Erlangen, Germany; 3Siemens Healthcare, Boston, United States; 4Harvard-MIT Division of Health Sciences and Technology, MIT, Cambridge, MA, United States

In this work we used a novel Connectom Gradient (AS302) system specifically designed to achieve ultra-high gradient strength (Gmax = 300 mT/m , about 7.5 fold stronger than clinical scanners). This enables to shorten the diffusion-encoding time and thereby decrease the TE, yielding significant gains in SNR. In addition, we developed a 64-channel receive coil to further increase the SNR. We compare the in vivo HARDI data with Q-Ball reconstruction at variable gradient strengths and b-values up to 15000 s/mm2 in human. The higher SNR is shown to provide improvements in ODF metrics such as the fiber uncertainty and FA.

Keywords

ability accuracy accurate achieve acquisition addition additionally allow allowing although angular appears applied architecture array assuming assured availability available ball better biomedical blueprint capable channel channels clinical coil commercially common considerable contribution corona corpus corrected custom dataset datasets decrease depict depicted designed detail developed development diffusion direct distribution division electrostatic enable encode encoding equipped every fiber fold function funded future gain gains general genuine gradient gradients harmonics health hospital human improved improvement improvements in vivo intersection interspersed isotropic limited long loss made matrix measures medical metrics middle models motion national noise orientation overlaid partial physiological posterior predicted preliminary prevent principle processing product progress providing ranging receive reception reconstructed reconstruction reduced reduction refocusing regularization remains repulsion resolution robust robustness scanned scanners school sciences secondary sharpening shorten slew slices society spacing specifically spherical stimulation strength strengths stronger substantial system systems technology tensors theoretical thereby third ultra uncertainty unclear variable volumes volunteer white yielded yielding