Optimal Flip Angle for Balanced SSFP Cardiac Cine Imaging

Subashini Srinivasan1, 2, Daniel B. Ennis, 13

1Department of Bioengineering, University of California, Los Angeles, CA, United States; 2Department of Radiological Sciences, University of California, Los Angeles, CA, United States; 3Department of Radiological Sciences, University of California Los Angeles, Los Angeles, CA, United States

Cine balanced SSFP is the clinical standard for cardiac structural and functional imaging due to its high blood-myocardium contrast compared to spoiled GRE imaging. The literature indicates that the optimal flip angle required for the maximum blood-myocardium contrast is 54˚ for imaging at 1.5T, but simulations didnt account for flowing blood. Our Bloch simulations of stationary myocardium and flowing blood indicate that a flip angle of ~130˚ is required for maximum blood-myocardium contrast and agree with in vivo results up to ~80. The highest available flip angle, subject to SAR limitations, should be used for balanced SSFP cardiac cine imaging.

Keywords

able achievable achieve agrees authors axis background balanced bioengineering blood cardiac chamber cine complex considering constant contrast currently density dependence depending depends determine diastolic drawn equation exceeding experimental experimentally explored five flow flowing grateful highest highlight identical imperfect importance increasing inflow likely limit limitation limited linear linearly literature measure measured medical myocardial myocardium objective optimal patterns percent preferred previous profile radiological radiology ranging regression relatively remains replacement resolution respect respectively sample scaled scanner sciences segmented short significantly simulated simulation simulations slice slices solutions spin spoiled stationary steady subject support theoretically traditional variable view views volunteer volunteers