Abstract #4427

Scanner Specific Relationship Between Displayed SAR and Heating Adjacent to Deep Brain Stimulation Leads: An in Vitro Study

Robert Dawe1, Leo Verhagen Metman1

1Rush University Medical Center, Chicago, IL, United States

The guideline for head specific absorption rate (SAR) during MRI of patients with implanted deep brain stimulation (DBS) leads is difficult to interpret because SAR estimation techniques vary among scanner manufacturers. The purpose of this study was to establish a relationship between displayed head SAR and local tissue heating in our particular scanner. Heating of approximately 4.9˚ C per 1.0 W/lb. of displayed head SAR was observed, confirming that potentially dangerous heating of DBS leads is possible. At the same time, the results suggest that the SAR guideline for implanted DBS leads may be overly conservative.

Keywords

acid actually added adjacent adult affixed align allowable always among apparently approximating arrangement audience bars basis blanket blood brain cause certainly close coil common confidence configuration conservative consistency constructed cooling curved damage deep described develop difficult displayed employed establishing estimating exactly exceeded expected explain exponential extension fiber free giving graph guideline hand head heating human immediately implant implanted in vivo indicate interval involved late lead leads least like limit limitation linear local maintain major manufacturer martin measure measures medical might mimicking mode model near neglecting neurologists neurosurgeons occur occurred occurring optic others overly particular patient patients perfusion permanent phantom physicists placed plotting probe produce produced protocols prudent published pulse pulses quality rage receive recorded relation relationship rely represent respective restrict rise routine rush safety saline scanner several severely shape short since slope software solid spin stimulation target technologists temperature thereby thoroughly thought threshold tissue torso transmit turbo type typical unilateral vertical vitro worthwhile