Integration of 2-Channel Parallel Transmission with Forced Current Excitation for Improved B1 Homogeneity in Breast Imaging at 7T

Sergey Cheshkov1, 2, Ivan E. Dimitrov1, 3, Wouter Koning4, Joseph Rispoli5, Mary P. McDougall5, 6, Steve Wright, 56, Craig R. Malloy1, 2

1Advanced Imaging Research Center, University of Texas Southwestern Medical Center, Dallas, TX, United States; 2Radiology, University of Texas Southwestern Medical Center, Dallas, TX, United States; 3Philips Medical Systems, Cleaveland, OH, United States; 4UMC Utrecht, Utrecht, CX, Netherlands; 5Biomedical Engineering, Texas A&M University, College Station, TX, United States; 6Electrical Engineering, Texas A&M University, College Station, TX, United States

Ultra-high field MR offers the promise of higher spatial/temporal resolutions in imaging, and improved sensitivity and quantification in MRS. Capitalizing on these advantages requires resolving the issues of decreased B1 homogeneity. We report on improved B1 homogeneity at 7T achieved by combining the flexibility of 2-channel parallel transmission with the insensitivity to loading of Forced Current Excitation breast coil. The overall transmit design is favorable for high quality breast imaging at high static magnetic field. Additionally, because of its insensitivity to coil loading this approach may allow for better patient throughput since breast size and composition vary widely across population.

Keywords

achieved acquisition adding additional additionally addressing advanced advantages allow amplifiers amplitude amplitudes aperture approaches approved better biomedical blue body breast calibration channel channels coil coils collected college combining common composition concomitant controlled decreased depicts deviation dispersion drawn driven either electrical element engineering ensure equal estimation even excitation exclude favorable feed field flatter flexibility forced freedom general hardware homogeneity horizontal human improved improvement improves independent indicating individual inhomogeneity insensitivity insufficient integration issue issues lead length loading local magnitude making maps medical middle minimization mode nominal notably offer offers optimization optimized overall pair parallel parametric patient phantom phantoms population power previously profile proposed protocol quad quadrature quantification radiology requires resolution resolutions resolving saddle scanner selected selective sensitivity setting several shim shimming simplified since slice southwestern spacing spatial spectral station studies subset systems temporal termed transmission transmit tuning twice unequal unilateral user utilizing variability variation vary vendor volume volunteer volunteers wave whole wright yellow