Calibration of Confounder-Corrected R2* for Liver Iron Quantification at 1.5T and 3T: Preliminary Results

Diego Hernando1, Naila Qazi1, Scott B. Reeder1, 2

1Radiology, University of Wisconsin-Madison, Madison, WI, United States; 2Medical Physics, University of Wisconsin-Madison, Madison, WI, United States

Measurement of liver iron concentration (LIC) is needed for detection and treatment monitoring of iron overload. R2-based techniques (eg: Ferriscan) are accurate at 1.5T, however they require long acquisition times. R2*-based techniques are fast, but suffer from several confounding factors: liver fat, background B0 variations and noise floor effects. In this work, we calibrated confounder-corrected R2*-based LIC quantification at 1.5T and 3T using multiple different protocols, with Ferriscan-LIC (1.5T) as reference standard. Nearly identical calibrations were obtained with different protocols at each field strength, suggesting that R2* may provide accurate and robust LIC quantification if relevant confounding factors are addressed.

Keywords

accelerator accordance accurate acknowledge acquisition addressed affect affected alumni artifacts audience avoiding axial background basic become bile blood breath calibrated calibration calibrations clinical closer complex concentration confounder confounding consistent construct controls coronal corrected correlation coverage curves detection determine discrepancy ducts either error essentially estimation excluding expected fact field fitting floor foundation free generally generated good half highest hold homogeneous identical intercept iron joint linear liver local long mapping maps medical monitoring nearly needed noise ongoing overload part physics preliminary presence previous program promising proposed protocol protocols quantification radiology rapid rapidly reeder regression relevant researchers respectively robust scanned segment sensitive several slices slope slopes spin staging strength studies subjects suffer suggests support target thank tissue treatment understood variability variations vessels water whether whole wish women wood