Abstract #4014

An Optimized Room-Temperature RF-Surface Resonator for in vivo Potassium-39 MRI at 9.4 T - Simulation and Measurement Study for Cryogenic Coils

Ibrahim A. Elabyad1, Raffi Kalayciyan1, Nagesh Shanbhag2, Lothar Schilling2, Lothar R. Schad1

1Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany; 2Division of Neurosurgical Research, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany

Potassium-39 (39K) Magnetic Resonance Imaging (MRI) is a non-invasive technique which could potentially allow for detecting intracellular physiological variations in common human pathologies such as stroke and cancer. However, the low signal-to-noise ratio (SNR) achieved in 39K-MR images hampered data acquisition with sufficiently-high spatial and temporal resolution in animal models so far. In order to improve the detector SNR gain, its optimal size, geometry, and temperature were determined via electromagnetic (EM) simulations and bench-top measurements. An optimized single-loop room-temperature surface resonator was then developed and tested for 39K-MRI of a healthy live rat brain. Compared to the previous published results of a triple resonant coil (1H, 23Na, and 39K), the developed resonator improve the SNR eight times with twice faster acquisition time. As experienced in the EM-simulations cryogenic-cooling of the optimal single-loop coil could further improve the SNR gain three-fold at 77 K and up to six-fold at 20 K. These significant results could further improve the available signal in future 39K-MR imaging studies of the rat brain at 9.4 T.

Keywords

achieve achieved acquisition advantageous analyzer appropriate approval approximately arranged assisted bench birdcage bottom brain built capacitor capacitors cerebral chemical clinical coherence coil coils column concentration concentrations conditions conjunction connected cooling copper coupling cryogenic cylindrical decay decreases developed diameter difficult dimensions double edge electro equations ethics evaluated experiment extracellular faculty fixed flash fold form frequency full gain gains geometries geometry half healthy hence herein house improve improved in vivo increasing indicating inductive inside insulating inter intracellular length linear live loaded local loop loss matched maximize measure measured medical medicine micro million monitoring much network noise occurs optimal orthogonally pathological potassium power previously properties prototype quantum reagents recent reflection resonant resonator resonators respectively return revealed room sample selected sensitivity significantly simulated simulation simulations since slice slices sodium spatially spiral steady stroke structures studies suffers superimposed surface system table temperature third tissue tuned tuning turn turns type unloaded validate validated vector vials view wave wright zero