Abstract #3848

Magnetic Susceptibility as a Field-Independent MRI Biomarker of Liver Iron Overload

Diego Hernando1, Scott B. Reeder1, 2

1Radiology, University of Wisconsin-Madison, Madison, WI, United States; 2Medical Physics, University of Wisconsin-Madison, Madison, WI, United States

MR-based quantification of liver magnetic susceptibility may enable field strength-independent quantification of liver iron concentration (LIC), for diagnosis, staging and treatment monitoring of hepatic iron overload. However, susceptibility measurement is challenging, due to the non-local effect of the susceptibility distribution on the measured B0. The purpose of this work is to demonstrate feasibility of MR-based LIC quantification using a fat-referenced approach to estimate liver susceptibility from the measured B0 field map. The proposed method is validated at 1.5T and 3T using an R2-based LIC measurement (Ferriscan) as the reference standard.

Keywords

abdomen abdominal ability accelerator acknowledge acquisitions adjacent agreement allows alumni approval aspect assigning audience automatically basic blood calibrated challenging chemical clinical closely complex concentration consent constitutes control controls correlated correlates correlation demonstrating density derived development diagnosis distribution dynamic enable encoded estimation example exploiting expressed feasibility feasible field foundation fraction good illustrate improve independent informed intercept internal iron known lateral liver lobe local mapping maps measured medical monitoring near needed organs overload phys physics potential potentially predicted presence program promising proton quantification quantify radiology reeder relationship relationships researchers scanned scientists segment separation simultaneous slope spin staging strength strongly subcutaneous subject subjects subsequently support susceptibility target thank theoretical tissue tissues treatment volunteers water wish