Abstract #3755

Bringing Prospective Motion Correction to Clinical Routine - Fast Camera Calibration

Cris Lovell-Smith1, Julian R. Maclaren2, Michael Herbst1, Ilja Kadashevich3, K.A. Danishad3, Oliver Speck3, Maxim Zaitsev1

1Department of Radiology, University Medical Center Freiburg, Freiburg, Germany; 2Department of Radiology, Stanford University, Stanford, CA, United States; 3Faculty of Natural Sciences, Institute of Experimental Physics (IEP), Magdeburg, Germany

Motion-induced artefacts in MR images degrade quality. The use of an optical tracking system to prospectively compensate for head movement during brain imaging has been shown to reduce these artefacts. To successfully apply motion correction, the cross-calibration matrix defining the transform between tracking system coordinates and scanner coordinates must be accurately known. We previously reported a method to determine this transform, however it required two experienced MR technicians upward of 20 minutes to perform. We present two newly developed methods that accurately calculate the cross-calibration within two minutes with one operator, thereby reducing the system setup time considerably. This is a major step in bringing prospective motion correction to clinical routine.

Keywords

accomplished accuracy accurately affine allow allowing altered angular another attached audience averaging become bore bottom bringing calculating calibration camera clinical compensate complete complex components comprises considerably constant coordinate coordinates correction cross damaged define defined degrees density designed determine edges either enabled equations error errors except execution experimental exploited facilitates fall fast fixed foreground forward freedom grid ground head improve improved initial initially innovation inside internal iterations iterative known major marker markers matrices matrix maxim measured minutes motion move moved movement must name named newly operator optical original patient performance phantom portability pose position positional previous procedure produced promising prospective prototyping quaternions question radiology random rapid reason recalibration recorded reduce reducing reduction registration retro returning root routine saved scanned scanner sciences section setting setup seven sharp short simplicity speed spherical square step stored structure sufficient surface symmetry system table target thereby tracking transform transformation transforms translation translational truth undergoes uniquely vectors volume