Evaluation of Readout-Segmented EPI for Use in fMRI at 7T

Ryan K. Robison1, 2, Allen T. Newton2, 3

1Philips Healthcare, Nashville, TN, United States; 2Vanderbilt University Institute of Imaging Science, Vanderbilt University, Nashville, TN, United States; 3Monroe Carell Jr. Children's Hospital at Vanderbilt, Nashville, TN, United States

Readout-segmented EPI is applied herein to functional MRI to reduce the effects of shot-to-shot signal variations. At ultra high field, geometric distortion may compromise the use of single-shot EPI for functional studies, particularly for high spatial resolution. Interleaved multi-shot EPI has been investigated as an alternative but is strongly affected by shot-to-shot signal variations. Improvement in temporal SNR is shown for readout-segmented EPI as compared to interleaved multi-shot EPI. Improved geometric distortion is also shown for readout-segmented EPI as compared to single-shot EPI.

Keywords

accommodates acquisition acquisitions alternative amplitude applying around artifacts audience background bandwidth best blurring bold children coherent concentrated considered consistent cost covers decay despite developments deviation distortion distortions dynamic dynamically encode encoding energy enough equivalent errors evaluation example expected experienced features fidelity functional furthermore geometric geometry ghosting ghosts gradient highest hospital improved improvement in vivo indicates indication inherent instability institute intended interleaved location locations loss magnitude majority make manifest maps masked methodological might mitigating much newton noise occur overwhelm patterns period periodic porter postulated produces producing proposed readout reconstructed reduced reduction resolution respiration scaled science segmented sensitivity series sharp shortest shot shots simulated sinusoidally slice smaller space stability static still strength strong studies suggested support suppress system taking target temporal temporally trade train unacceptable variations varied varying volume white yellow zero