Meeting Banner
Abstract #3132

Exact Analytical Results for ADC for Oscillating Diffusion Sensitizing Gradients

Alexander L. Sukstanskii1

1Radiology, Washington University, St. Louis, MO, United States

Apparent diffusion coefficient (ADC) is analyzed for the case of oscillating diffusion sensitizing gradients. The exact analytical expressions are obtained for the high-frequency expansion of ADC for an arbitrary number of oscillations. The validity conditions of the high-frequency expansion of ADC are analyzed. It is shown that these conditions are substantially different for cos- and sin-type gradients. For the cos-type gradients, the high-frequency regime is reached when T<<tD (T is the period of oscillations, tD is the characteristic diffusion time), whereas for the sin-type gradients, this regime is achieved when the total diffusion time t is smaller tD.

Keywords

achieve amplitude analytical analyze analyzed apparent apply arbitrary available basing become behavior biological boundaries calculating challenging characteristic chem close closed coefficient coincides communication condition conditions consider considered consists contrast convenient correspond course dependence depends derived described diffusion dimensionality divergence dynamic enough equal exact exactly expansion explore expression expressions form free frequency full function functions generally geometry gradient gradients hand impermeable importantly imposes independent intermediate interval leads length limit limiting measuring media model much namely narrow never numerically oscillating oscillation oscillations panel parsons period periods phys porous practically problem promising proportional pulse radiology regime reliable reminded require respectively restricted restriction restrictions rewrite scales segment sensitizing short simple smaller softer solution speaking substantially substituted sufficient sufficiently surface system systems technically type usually valid validity varies volume whereas