Meeting Banner
Abstract #2826

RF-Heating Testing in 64 MHz RF-Laboratory System and 1.5 Tesla MRI a Comparative Evaluation

Wolfgang Grtz1, Nicolas Flle1, Gerrit Schnwald1, Susanne Matthey1, Gregor Schaefers1

1MR:comp GmbH, Germany, Gelsenkirchen, NRW, Germany

RF-heating tests as described in ASTM F2182 have been performed in a 1.5 Tesla MR scanner and additionally in a 64 MHz RF laboratory workbench system. A comparison of the results of the detected temperature increases shows a good agreement between both systems. Based on this finding it can be concluded that the laboratory system is an alternative to usually used clinical MR systems for RF heating test on medical implants following ASTM F2182.

Keywords

according accuracy acrylic alternative application applying around avoid band bandwidth calculations calibration clinical close comp comparative computer configurations confirmed conformity contra control cranial cycle described detectable detected determined dimensions displayed distribution done duration duty ends environment evaluation expected exposure fiber field finding fluid forecast frequency glass good gradient greater greatest heating height implant implants important increasing induced inner laboratory lateral leads located magnet maximal medical middle minutes mixed mode must nearly numerical object onto optic overall part phantom pixel placed position positioned positioning positions power prior probe probes pulse radio refocusing related replace required resolution resonant risks runs scanning settings shot side simulation software spacing start stop suitable system systems target temperature thermometer titanium type verify visible whereas whole workbench worst