Meeting Banner
Abstract #2716

Analysis of Measurement Precision in Continuous Magnetic Field Monitoring

Benjamin E. Dietrich1, David Otto Brunner1, Bertram J. Wilm1, Christoph Barmet1, 2, Klaas P. Pruessmann1

1Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland; 2Skope Magnetic Resonance Technologies, Zurich, Switzerland

Dynamic magnetic field monitoring with NMR probes enables the observation of the spatio-temporal magnetic field evolution during MR experiments. A recently proposed method based on time interleaved acquisition of sets of fast relaxing NMR probes allows scanner independent dynamic magnetic field monitoring even under strong gradients over arbitrary durations. Such continuous gradient field monitoring alleviates hence the limitation on resolution, acquisition duration and duty cycle of the approaches shown using a single coherence. The presented work assesses the precision of this method with regard to different field-probe-set configurations, gradient strengths, set alternation patterns, and re-excitation periods.

Keywords

achieved acquisition actually allowed allows alone alternation always amplitude applied approaches arranged assumption avoided bandwidth bars behave biomedical black built camera capillaries chains coherences coils concluded concomitant conditioned configuration configurations consisted constant continuous contribute contributions converters copper corners course darker decay degrees delivers demodulation dependence deviation diameter diameters digital doped duration dynamic enables error errors even evolution excitation excited expect expected fact fast favorable field fields filtered finally findings formation gaps glass gradient gradients head hence induced initial inner inset instruments interleaved interpolated interpolation interval involved just larger length lies limitation limited linear long made major materials measure measured measures mechanism monitoring mounting moved noise observation overall part period periods phasing plates positions precision principle probe probes radial receive regard residual selected sensitivity served sets setup several severe shorter solenoids source sources spacing spurious stand strength strengths strictly strong system taken temporal terms tetrahedron thermal thread transmit typical violation volume wire