Abstract #2598

Rapid B1 Mapping Method for Multi-Channel RF Transmit Coil Using Phase-Difference

Kosuke Ito1, Masahiro Takizawa1, Tetsuhiko Takahashi1

1MRI system division, Hitachi Medical corporation, Kashiwa, Chiba, Japan

Fast B1 mapping method for multi channel transmit coil is developed. In this method, B1 map acquisition is only one time with using all transmit channels. The acquired B1 map is decomposed to B1 map for each transmit channel by using the information of phase difference between transmit channels. Without repeating B1 map acquisition, short scan time is achieved. Also, only B1 map for all channel transmission is used, this method relaxes the required accuracy of B1 map. For 4-channel transmit coil, scan time is only 1 sec.

Keywords

abstract accuracy achieved acquisition advantage another applications applied available body calculation changing channel channels clinical coil combine combined corporation dashed decomposed delay denoted determined developed diagonal discovered discuss dividing division extension extent fast field filtering general gradient improve increasing japan known like made make mapping maps mask materials matrix measured medical multiplied needed parallelogram phantom practice procedure proposed pulse quantity quite rapid receive received recently reconstruction relation relaxed repeat repeatedly repeating require required routine sense shield shimming short simple simplicity since slice smoothing solid spatial straightforward system systems theory threshold torso transmission transmit transmitted triangle vector whole