Abstract #2595

SNR Analysis of Adiabatic Bloch-Sigert B1+ Mapping

Mohammad Mehdi Khalighi1, Jason H. Su2, 3, Brian K. Rutt3

1Applied Science Lab, GE Healthcare, Menlo Park, CA, United States; 2Department of Electrical Engineering, Stanford University, Stanford, CA, United States; 3Department of Radiology, Stanford University, Stanford, CA, United States

Adiabatic Bloch-Siegert (ABS) B1+ mapping with spiral readout has been recently introduced as a fast and efficient way of volumetric B1+ mapping with no T1 or T2 dependency. Here we have analyzed the angle-to-noise ratio (ANR) of the ABS method and show phantom and brain ANR maps at 7T. We also compare analytically the phase-based ABS using spiral readout with the magnitude-based DREAM method. Our comparison shows that ABS generates >3 times ANR compared to DREAM for similar scan times.

Keywords

achieved addresses adiabatic allows amplitude amplitudes amps analytical applications applied approved array assume assumed assuming bandwidth benefits blue brain candidates channel channels choose coil comparable computed constant conversion depends deposition derived determine deviation distribution dots dream effective electrical engineering equation estimator excitation expected experiment expression fast faster fold frequency gain generate generated generates gives good gradient grant greater head highest in vivo introduced introducing investigate last limits long makes mapping maps matched matrix measure measured measuring medical noise nova offer optimal others parallel park phantom power predicts prescribed problems pulse pulses quadratically quality radiology rather readout receive recently repeated resolution rows scanner science sensitivity series short shorter shot silicon similarly slice slices spiral support swept theory transmit turn uncorrelated variance volumetric volunteer whole written