A T2-Diffusion-Prepared Cube Sequence for Brain Lesion Detection at 7T

Eric K. Gibbons1, Manoj Saranathan2, Brian K. Rutt2, John M. Pauly3, Adam B. Kerr3

1Department of Bioengineering, Stanford University, Stanford, CA, United States; 2Department of Radiology, Stanford University, Stanford, CA, United States; 3Department of Electrical Engineering, Stanford University, Stanford, CA, United States

Brain lesion detection at 7T is complicated by the T1 lengthening observed for white and grey matter without a corresponding increase in CSF T1. As a consequence, there is lower SNR and contrast for grey and white matter in FLAIR sequences designed to null CSF signal. The recently developed MP-FLAIR 3D FSE sequence introduced a magnetization preparation sequence to initially saturate white/grey matter Mz so as to allow for greater recovery during the FLAIR inversion period. In this preliminary work, we examine the possibility of eliminating the inversion preparation entirely, using instead a T2 and diffusion preparation sequence, with the goal being CSF signal suppression due to its higher diffusion, and simultaneous introduction of combined T2/diffusion contrast between white/grey matter and lesions.

Keywords

accurately acquisition adiabatic adjusted affecting allow alternative amplitude anisotropic apply arise artifacts axes axis bioengineering bipolar blurring brain cardiac central channel coil combined complicated consequence contrast coronal cube dark decay delay dependence depending designed desired despite detection developed diffusion done dramatic drop echoes eddy effective electrical eliminating encodes engineering entirely examine excitation final flair flow frequency future gating ghosting gibbons goal gradient gradients greater head healthy hemisphere implemented improving in vivo incorporated increasing inhomogeneities inhomogeneity initially instead intervals introduced inversion isotropic iterative john lengthening lesion lesions limit lobe lobes loss magnetization male manifested minimize mitigate module motion navigators needs novel null observe occur optimized partly period positive possibility preliminary preparation prepared produced pseudo published pulse pulses radial radiology recently reconstruction recovery reduce reducing refocused restores sampling saturate scanner scanning selective sensitivity shorten simulation simulations simultaneous simultaneously slew slice space spatial supported suppression surrounding tissues transverse trigger twice uniform varied verified volunteer white widths