Calibration of Dynamic Susceptibility Contrast MRI Using T1-Based Steady-State CBV (Bookend Technique) and Vascular Space Occupancy (VASO): Comparison with Pseudo-Continuous Arterial Spin Labeling

Emelie Lindgren1, Ronnie Wirestam1, Karin Markenroth Bloch1, 2, Andr Ahlgren1, Matthias J.P. van Osch3, Danielle van Westen4, Yulia Surova5, 6, Freddy Sthlberg1, 7, Linda Knutsson1

1Department of Medical Radiation Physics, Lund University, Lund, Sweden; 2Clinical science, Philips, Lund, Sweden; 3C.J.Gorter Center for high field MRI, Department of Radiology, LUMC, Leiden, Netherlands; 4Center for Medical Imaging and Physiology, Skane University Hospital Lund, Lund, Sweden; 5Department of Clinical Sciences, Lund University, Lund, Sweden; 6Department of Neurology, Skne University Hospital, Lund, Sweden; 7Lund University Bioimaging Centrum, Lund University, Lund, Sweden

Quantification of CBF in absolute terms is relevant in some clinical applications. However, absolute CBF obtained by dynamic susceptibility contrast MRI (DSC-MRI) is usually overestimated. One solution is to calibrate DSC-MRI CBF using a complementary CBV method. In this study, T1-based CBV (Bookend) and VASO were used to calibrate CBF from DSC-MRI, and results from calibrations were compared with CBF from pseudo-continuous ASL (pCASL). Both calibration methods showed absolute grey matter CBF values which were comparable with corresponding CBF values from pCASL.

Keywords

absolute according acquisition administration agent although another applications applied approaches arterial bland blood bolus bookend calibrate calibrated calibration called cerebral certain clinical collected commonly comparable comparing complementary considerably continuous contrast correspond curve days delay done duration dynamic employed estimation example executed expected extracted field fitting flow green healthy hospital importance improve included increment independent indicates injection inversion investigations label labeling leading likely locker look magma maps material matrix measured measuring medical motion need neurology nulls occupancy often optimized overestimated partial percent perfusion physics physiology plot plots post potential potentially problematic pseudo quantify quite radiation radiology reasonable recorded recovery relevant rely repetition require resolution respect respectively retest return scanned science sciences segmentation segmented selective shin slice slices slightly smaller space spin steady subject subjects susceptibility system temporal tended terms thicknesses twenty twice uncalibrated underestimate unfortunately unusable useful vascular volume volumes volunteers white whole