ASL Inversion Efficiency for Three Methods and Two Magnetic Fields

Clment S. Debacker1, 2, Jan M. Warnking1, 3, Jrme Voiron2, Emmanuel Luc Barbier, 34

1Team 5: Functional NeuroImaging and Brain Perfusion, Grenoble-Institute of Neuroscience, La Tronche, France; 2MRI, Bruker BioSpin, Ettlingen, Germany; 3U836, INSERM, La Tronche, France; 4Team 5: Functional NeuroImaging and Brain Perfusion, Grenoble Institute of Neurosciences, La Tronche, France

Arterial Spin Labeling (ASL) methods can quantify cerebral blood flow (CBF), but this requires an estimate of the inversion efficiency (IE). The purpose of this study is to estimate IE for three ASL methods (pulsed ASL (PASL), continuous ASL (CASL), and pseudo-continuous ASL (pCASL)) and two magnetic fields in the context of pre-clinical imaging. We also show that shim correction seems to increase IE.

Keywords

acquisition adiabatic agreement amplitude appears around arterial assertion audience bandwidth best blood brain capable carotid cerebral check chosen close coil comparable compensated computed conditions conducted configuration considerations constant context continuous control correct correction decrease decreases degree delay despite determination duration efficiency equilibrium error errors expected experiment explained field fields flow functional generating global good gradient homogeneity horizontal human hyperbolic imply in vivo inhomogeneity institute inversion inversions label labeling larger literature longitudinal magnetization male matrix measure measured methodologists might modalities module noninvasive perfusion preliminary profiles promising pseudo pulse pulsed quantify rats reach realized receive reception reduced reduction reproducibility required requires residual rise scanners secant seem selective sensitive sensitivity shape shim similarly spin strategy studies surface tagging target team tissue transit transmit unpaired verify volume water