Simultaneous Multislice Readout-Segmented Diffusion-Weighted EPI with Blipped-Controlled Aliasing

Samantha J. Holdsworth1, Rafael O'Halloran1, Anh Tu Van1, Eric Aboussouan1, William A. Grissom2, Anuj Sharma2, Murat Aksoy, Julian R. Maclaren, Stefan Skare3, Roland Bammer

1 (Equal contribution): Center for Quantitative Neuroimaging, Department of Radiology, Stanford University, Palo Alto, CA, United States; 2 (Equal contribution): Biomedical Engineering, Vanderbilt University, Nashville, TN, United States; 3Clinical Neuroscience, Karolinska Institute, Stockholm, Sweden

Readout-segmented EPI (RS-EPI) has been shown as a promising candidate for reducing distortion in diffusion-weighted (DW)-EPI while being robust to motion-induced phase errors. However, the requirement of several adjacent segments (or 'blinds') in RS-EPI can make the scan time can prohibitively long particularly for thin slices which require a large number of slices to achieve full brain coverage. In this work, we reduce the minimum TR in RS-EPI with the use a simultaneous multislice acquisition using PINS multiband pulses coupled with blipped-controlled aliasing.

Keywords

accelerate acceleration accompanied achieve achieved acquiring acquisition acquisitions adjacent advanced aliased aliasing alternating alto appear application applications applied approaches audience band bands better biggest biomedical blind blinds blipped blips brain calibration candidate channel chemical clinical coil column combine concerns contribution controlled correction coupled coverage designed diagram diffusion distance distortion effectively efficiency efficient encoding engineering entropy equal errors essentially estimation excitation excite excited expected exploit extra faster feasible fidelity final fold form foundation full geometric ghost gradients gridded hard head healthy individually induced inherent institute limited long make matrix motion moved navigator need negative note opposite particularly pins polarity positive preparation prohibitively promising providing pulse pulses quantitative radiology readout readouts reconstruction reduce reducing refocus refocused refocusing removing replacing require resolution saturation segmented segments select selected selects sensitivity separated separation series several simultaneous simultaneously slice slices space spaced steps suppression suppressive system tanner target technology thin timing together trajectory underwent upfront volunteer width